Difference between revisions of "TODO"

From BESIII Ferrara Group Wiki
Jump to: navigation, search
m
m (Possible solutions to be confirmed)
Line 21: Line 21:
 
== Possible solutions to be confirmed ==
 
== Possible solutions to be confirmed ==
  
=== The charge measured with random trigger is above 10fC, up to 50 fC ===
+
==== The charge measured with random trigger is above 10fC, up to 50 fC ====
 
This is due to few channels with very high noise:
 
This is due to few channels with very high noise:
 
* FEB_label = 21, chip 2, channel 3
 
* FEB_label = 21, chip 2, channel 3
* FEB_label = 23, chip ?, channel 26
+
* FEB_label = 23, chip 2, channel 26
 
* FEB_label = 36, chip 2, channel 3
 
* FEB_label = 36, chip 2, channel 3
 
* FEB_label = 43, chip 2, channel 45
 
* FEB_label = 43, chip 2, channel 45
  
 +
3/4 of these channels (ch. 3 and 45) are located on the edge of the anode-FEB connector (see [[L2_FEB_pinout]]).
  
 +
TO-DO:
 +
* Check and confirm that removing these channels the very high charge noise disappears (for all runs)
 +
* Check the FEB-anode connection
 +
* Check the position of these FEBs --> are they close to some potential noise source?
  
 
== Solved ==
 
== Solved ==
  
 
* No signal from FEB 2 and 30 --> due to missing calibration files (raw data should be present)
 
* No signal from FEB 2 and 30 --> due to missing calibration files (raw data should be present)

Revision as of 01:56, 28 May 2019

List of bug and problems[edit | edit source]

  • Charge decrease as the threshold increases -> check noisy channel
  • GemHit_time shows some holes -> check if it is fixed with tfine
  • The reconstruction is too slow with event.C -> limited the reconstruction to the first 10k events
  • The radius used on the L2 do not correspond to the real on
  • µTPC disabled up to now
  • Only 33% of the hits has a cluster on both sides of the L2, while in principle it should be 100% --> problem due to ROC?
  • FEB from 0 to 15 are noisy at channel 62
  • Peaking background on FEB=6 && channel = 62


Possible solutions to be confirmed[edit | edit source]

The charge measured with random trigger is above 10fC, up to 50 fC[edit | edit source]

This is due to few channels with very high noise:

  • FEB_label = 21, chip 2, channel 3
  • FEB_label = 23, chip 2, channel 26
  • FEB_label = 36, chip 2, channel 3
  • FEB_label = 43, chip 2, channel 45

3/4 of these channels (ch. 3 and 45) are located on the edge of the anode-FEB connector (see L2_FEB_pinout).

TO-DO:

  • Check and confirm that removing these channels the very high charge noise disappears (for all runs)
  • Check the FEB-anode connection
  • Check the position of these FEBs --> are they close to some potential noise source?

Solved[edit | edit source]

  • No signal from FEB 2 and 30 --> due to missing calibration files (raw data should be present)